Friday, February 24, 2012

Methods: clinical case reports of 56-year ...

Objectives: To report a case of a large subcutaneous emphysema introduced during vitrectomy using advanced feedback pressure infusion system. Methods: Clinical case report 56-year-old woman undergoing pars plan vitrectomy for rhegmatogenous retinal detachment left eye. Clinical and radiological results of eye patients have been reported. Mechanisms for feedback pressure infusion devices have been considered to explain the side effects. Search PubMed search was conducted of any similar incidents and / or discussion. Results: In this case surgery, vitrectomy was completed with air fluid exchange and formed anterior chamber there was an estimated pressure high teens low 20's to the surgeon. After undraping, the patient was noted with serious facial crepitus apply to the clavicle. Immediate chest X-ray was done, and then use CT to confirm the orbital and subcutaneous emphysema, and the presence of perfluoro-N-octane in the left orbit. Conclusion: Integrated pressure infusion devices using feedback sensors can support complex method of intraocular pressure and formation of the globe. However, accidental or occult world gap can lead to disruption management with feedback and subsequent high rates of infusion. How intertwined substances leave the spot break, they can lead to large external gas or fluid accumulation, and we lasix 12 mg report a case of severe bilateral subcutaneous emphysema as a result. .